ABX Pentra 60 C+

ABX Pentra 60 C+ Haematology Analyzer

 

Overview

  • Data management on stand-alone PC.
  • Micro-sampling from whole blood (CBC : 30 µL – DIFF: 53 µL)
  • Cytochemistry, Impedance (real cell volume measurement) & Optical (analysis of the internal cellular structure by measuring light absorbance) – DHSS* Technology.
  • Perfect homogenization of blood samples with reagents – MDSS** Technology.

Features

  • Throughput: Up to 60 samples/hour
  • Reagents: Only 4 onboard reagents and 1 diluent
  • Perfect differentiation of the 5 WBC sub-populations with DHSS* Technology
  • 3 histograms for RBC, BAS/WBC and PLT together with the 5 DIFF Matrix.
  • Basophils counted through specific channel
  • High resolution matrix includes the determination of 2 additional subpopulations (% and #): Atypical Lymphocytes (ALY***) and Large Immature Cells (LIC***)

 

* DHSS : Double Hydrodynamic Sequential System 

** MDSS : Multi-Distribution Sampling System

*** RUO parameters (Research Use Only)

    Our package can include some or all of the following.

     

    For information on leasing me, Click Here.

     

    Product Selection Ensuring we get you the right product at the best price
    Installation and Debris Removal Ensuring the installation is complete to our standards and we recycle all packaging material
    Calibration / Preventative Maintenance Ensuring the equipment is maintained and/or calibrated as per ISO17025 standards or equivalent (certificate provided after installation)
    Extended Warranty with Preventative Maintenance Ensuring you are always covered no matter what and when equipment failures may occur
    Technical
    Weight 77 lbs / 35 kg
    Power supply 100 to 240 V (± 10%) – 50 to 60 Hz.
    Dimensions 20.3 x 17.5 x 19 in / 51,5 cm x 44,5 cm x 48 cm
    Printer Laser printer
    Throughput 60 tests/hour.
    Screen 16 lines LCD screen.
    Data management Color screen: 15 in. monitor (800 x 600px)

    Windows XP Prof.TM

    User-defined flagging limits

    Transmit patient & QC to LIS

    Uni-directional or bi-directional connection

    Communication Protocols: ABX / ASTM

    CD-ROM drive / Floppy disk

    2 serial ports minimum

    USB Keyboard / USB Mouse

    Memory 10,000 results + Graphics
    Quality Control Management 12 selectable QC files

    XB: 60 Operator selectable batches with statistics (20 samples per batch)

    Within run

    Levey-Jennings Graphs

    Logs Reagents, calibration, maintenance, errors, blank cycle
    Samples
    60 samples/hour (closed tube)
    Parameters (26) WBC, NEU# & NEU%, LYM# & LYM%, MON# & MON%, EOS# & EOS%, BAS# & BAS%, ALY# & ALY%*, LIC# & LIC%*, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT*, PDW*.
    Graphs High-definition LMNE matrix
    PLT, RBC, WBC/BAS curves.
    Blood Volume CBC mode : 30 µL
    CBC + DIFF mode : 53 µL
    Stability 48 hours post-draw stability
    Operating sample mode Open tube
    Identification Alphanumeric. Bar-code reader (optional)
     * RUO (Research Use Only)

     

     Results
    Parameters Precision (%CV)
    Linearity
    WBC < 2.0 0-120 x 10³/µL
    RBC < 2.0 0-8 x 106/µL
    HGB < 1.0 0-24 g/dl
    HCT < 2.0 0-67%
    PLT (whole blood) < 5.0 0-1900 x 10³/µL
    PLT (PLT concentrate) < 5.0 0-2800 x 10³/µL
    Blood Controls & Calibrators
    Blood Controls
    Reference Description Dimensions
    2062207 ABX Difftrol – 2L + floppy 2 x 3 mL vial
    2062203 ABX Difftrol – 2N + floppy 2 x 3 mL vial
    2062208 ABX Difftrol – 2H + floppy 2 x 3 mL vial
    2042208 ABX Minotrol 16 – 2L 2 x 2.5 mL vial
    2042209 ABX Minotrol 16 – 2H 2 x 2.5 mL vial
    2042202 ABX Minotrol 16 – 2N 2 x 2.5 mL vial
    Calibrators
    Reference Description Dimensions
    2032002 ABX Minocal 1×2 mL

    Reagents and Controls

    Haematology Reagents and Controls Click Here